• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22302D
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem Foreign Body In Patient (2687)
Event Date 07/04/2017
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to the manufacturer for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection of the bladder tumor (turbt) procedure, the loop wire at the distal end of the hf resection electrode broke off and fell inside the patient's bladder.However, no fragment remained inside the patient since it was reportedly retrieved.The intended procedure was successfully completed with another hf resection electrode and there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was returned to the manufacturer for evaluation/investigation.The evaluation/investigation confirmed that the hf resection electrode is damaged.The loop wire at the distal end is broken off completely.However, no part is missing since the fragment was returned as well.Furthermore, the hf resection electrode is deformed.The cause of this damage and the breakage of the loop wire is mechanical overload by the application of excessive force.Therefore, this event/incident was attributed to use error.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed from olympus side with no further actions.However, the reported event/incident will be recorded for trending and surveillance purposes.In addition, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 12°, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 
4940669662
MDR Report Key6750061
MDR Text Key81500036
Report Number9610773-2017-00096
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK100275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model NumberWA22302D
Device Catalogue NumberWA22302D
Device Lot Number16120P01L002
Other Device ID Number14042761051665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/07/2017
Initial Date FDA Received07/28/2017
Supplement Dates Manufacturer Received09/29/2017
Supplement Dates FDA Received10/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-