SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 397002-001
Device Problems Battery Problem (2885); Device Operates Differently Than Expected (2913); Device Sensing Problem (2917)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver would not recognize external batteries.
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the companion 2 driver would not recognize external batteries.
 
Manufacturer Narrative
The companion 2 driver was returned to syncardia for evaluation.A review of the patient data file alarm history confirmed the customer-reported issue, although it could not be reproduced or verified through investigation testing.Additionally, the external batteries that were reportedly used with the driver were not returned and therefore not evaluated or included in this investigation.The driver passed all functional testing requirements.Additionally, external batteries that have previously met performance requirements were used to test the ability of the driver to recognize and charge external batteries, which the driver was able to do.There was no evidence of a device malfunction and the driver performed as intended.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key6753825
MDR Text Key81532056
Report Number3003761017-2017-00133
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2017
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/20/2017
Initial Date FDA Received07/31/2017
Supplement Dates Manufacturer Received04/20/2018
Supplement Dates FDA Received05/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-