Model Number 595000-001
Device Problems Difficult to Remove (1528); Sticking (1597)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 07/18/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom onboard batteries were stuck in the driver and had to be pried out.
 
Manufacturer Narrative
The freedom driver was returned to syncardia for evaluation.The customer-reported battery extraction difficulty was confirmed and reproduced during investigation testing.Visual inspection of the driver found extensive damage to the left battery well and button and a broken right safety battery latch, indicating rough handling/improper use.The broken right safety battery latch did not allow the left onboard battery to be removed by hand from the driver during the onboard battery insertion/extraction test, confirming the customer-reported issue.The root cause of the onboard battery not being able to be removed from the driver is the broken safety battery latch.The right safety battery latch was found to be damaged, most likely caused by rough insertion of the batteries into the driver.A broken safety battery latch may prevent a battery from being extracted from the driver or from being fully latched into place.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom onboard batteries were stuck in the driver and had to be pried out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6753864
MDR Text Key81610021
Report Number3003761017-2017-00136
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/18/2017
Initial Date FDA Received07/31/2017
Supplement Dates Manufacturer Received08/30/2017
Supplement Dates FDA Received09/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-