• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Model Number CUSTOM COMBI SET
Device Problems Fluid/Blood Leak (1250); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/08/2017
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
The clinic manager at a user facility reported to post market surveillance that the arterial line of the bloodline dislodged from the arterial port of the central venous catheter (cvc) during treatment.There was no harm or adverse event reported.No parts were available to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
The device was not returned to the manufacturer for physical evaluation, and the lot number was not provided.Therefore, the failure mode cannot be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.A manufacturing review was performed of the products shipped to the dialysis center for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius bloodline products from the reported catalog number (03-2722-9) shipped to this account within the selected time frame.A records review was performed on all lots identified.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances identified during the manufacturing process which could be associated with the reported event.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.The lots passed all release criteria.A review of the batch production records did not reveal a probable cause for the customer complaint.There were no non-conformances identified that relate to the reported event, and all lots met release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6760285
MDR Text Key81609954
Report Number8030665-2017-00500
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Model NumberCUSTOM COMBI SET
Device Catalogue Number03-2722-9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Initial Date Manufacturer Received 07/12/2017
Initial Date FDA Received08/02/2017
Supplement Dates Manufacturer Received08/03/2017
Supplement Dates FDA Received08/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-