• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problems Expulsion (2933); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/18/2017
Event Type  malfunction  
Manufacturer Narrative
The complaint is being reported by zimmer biomet as (b)(4).(b)(4).The product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that before surgery the battery had exploded in the packaging.There was black dust in the packaging.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This follow-up report is being submitted to relay additional information.Review of the device history record identified no relevant deviations or anomalies.Product examination found that the unit had ruptured inside the sterile packaging before use.This complaint is confirmed.The root cause of the reported event is a short circuit of the battery.The source of the short could not be identified as the damage caused by the rupture was too extensive.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key6765096
MDR Text Key81852312
Report Number0001526350-2017-00453
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Model NumberN/A
Device Catalogue Number00515048200
Device Lot NumberZ000006612
Other Device ID Number(01)00889024375185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-