The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.In addition, damage on the antenna coil and shield wire was observed.The photographic imaging inspection confirmed antenna coil and shield wire breaks.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The residual gas analysis result revealed nitrogen levels above the limit, indicating a non-hermetic device.Due to the presence of moisture getter, no signs of moisture were observed.The reported complaint of loss of lock was verified during this analysis.This device had broken antenna coil wires, which was determined to have caused the loss of lock.Advanced bionics will continue to monitor failure modes of this type.In addition, based on dye penetrant data, it is determined that this device was non-hermetic.A corrective action was implemented.This is the final report.
|