Catalog Number 5MAXACE068KIT |
Device Problems
Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311); Device Handling Problem (3265)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the penumbra system ace 68 reperfusion catheter (ace68) was ovalized approximately 97.0-117.0 cm distal from the hub.The ace68 was kinked approximately 131.0 from the distal tip.Conclusion: evaluation of the returned ace68 confirmed that the distal tip was kinked and there was ovalization along the catheter body.This type of damage can occur if the ace68 is forcefully retracted at extreme angles from its product tray.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
During preparation for a thrombectomy procedure, the hospital staff noticed that the tip of the penumbra system ace 68 reperfusion catheter (ace68) was pinched upon removal from the packaging of the penumbra system ace 68 kit (kit).The damaged ace68 was found prior to use and therefore, was not used in the procedure.The procedure was completed using a new kit.
|
|
Search Alerts/Recalls
|