It was reported that shortly after a bolus, the patient encountered an occlusion alarm (alarm number in pump history: 26) during use of a cleo® 90 infusion set.Through troubleshooting with the distributor, it was determined that the blockage was likely located at the site.The patient's blood glucose levels were 247mg/dl at the time of the event.To address the high blood glucose levels, the patient was going to deliver a bolus and "had backup if necessary".The patient did not test for ketones.No permanent injury was reported.
|