Model Number GIF-H180J
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2017
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to omsc but was returned to olympus (b)(4).(b)(4) sent the subject device to a third party laboratory for additional microbiological testing.As a result of additional microbiological testing by a third party laboratory, no microorganisms were detected from samples taken from the instrument channel, auxiliary water channel and air / water channel of the subject device, but 3cfu microorganisms were detected from suction channel.The test result satisfied the french guideline.Also, omsc reviewed the manufacturing history of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If significant additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that in the routine microbiological test by the user facilities, microorganisms (more than 26 cfu/100ml) were detected from samples taken from the subject device.The type and number of microorganisms are unknown.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is submitting to correct "device product code".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key6772846
MDR Text Key82156178
Report Number8010047-2017-01171
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK051645
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H180J
Other Device ID Number04953170339110
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/19/2017
Initial Date FDA Received08/07/2017
Supplement Dates Manufacturer Received02/13/2019
Supplement Dates FDA Received03/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-