ZIMMER SURGICAL, INC. PULSAVAC PLUS FAN SPRAY KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problem Expulsion (2933)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/19/2017
Event Type  malfunction  
Manufacturer Narrative
The complaint is being reported by zimmer biomet as (b)(4).The product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during preparation for surgery the product looked like it had exploded inside of the sterile package.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the device history record identified no relevant deviations or anomalies.Product examination found that the battery pack had ruptured from overheating batteries inside the sealed sterile packaging.This complaint is confirmed.The reported event claimed that the sealed unit had ruptured before it had been opened.While rare, this kind of event occurs when the batteries overheat from a short circuit.To address this issue, a change notice has been implemented.The root cause of the reported event could not be specifically determined with the provided information.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSAVAC PLUS FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key6789704
MDR Text Key82708975
Report Number0001526350-2017-00495
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot NumberZ000007132
Other Device ID Number(01)00889024375161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-