Model Number MMT-723NAS
Device Problems Device Displays Incorrect Message (2591); Electromagnetic Compatibility Problem (2927)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/20/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call the insulin pump alarmed motor error and a motor position encoder error.The customer¿s blood glucose level was unknown at the time of the incident.The customer stated that the insulin pump had multiple motor error alarms.The customer stated that the drive support cap was normal and that the insulin pump was exposed to high magnetic fields.Customer also reported to have received a motor position encoder error.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Manufacturer Narrative
Motor error alarm during rewind due to motor encoder signal out of phase.Unable to perform self test, off no power test, unexpected restart error test, occlusion test, prime test, excessive no delivery test, displacement test due to motor error alarm.Pump passed idle current test and run current test.Unable to verify motor position encoder error alarm due to motor error alarm.Pump received with cracked display window, cracked battery tube threads and cracked reservoir tube lip.
 
Manufacturer Narrative
The correct information has been provided with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6790124
MDR Text Key82938410
Report Number2032227-2017-36322
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169507258
UDI-Public(01)00643169507258
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date 10/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Device Lot NumberB5723NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/20/2017
Initial Date FDA Received08/11/2017
Supplement Dates Manufacturer Received07/20/2017
10/07/2017
Supplement Dates FDA Received08/28/2017
10/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age15 YR
Patient Weight170
-
-