• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL, INC P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number MICROBLENDER
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 06/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Any additional information received from the customer will be included in a follow-up report.Results of investigation: the suspect device was returned to vyaire's factory service for evaluation and repair.The factory service representative was unable to duplicate the customer's reported event of the alarm issue but reported the front control knob was loose.After tightening/re-calibrating the front control knob, the issue was resolved.The factory service representative performed testing and reported the device meets service specifications.
 
Event Description
The customer reported while using the p/n 9320 low flow air/oxygen microblender; the device constantly alarms despite being plugged in to both air and oxygen.During evaluation of the suspect device, the vyaire factory service representative reported the front control knob is low as the measured fraction of inspired oxygen (fio2) is 51% at all settings.There was no patient involvement regarding this reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6793363
MDR Text Key82774536
Report Number2021710-2017-06467
Device Sequence Number1
Product Code BZR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMICROBLENDER
Device Catalogue Number03920A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/18/2017
Initial Date FDA Received08/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-