• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS MIH MOD CUP INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS MIH MOD CUP INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Separation Failure (2547)
Patient Problem No Information (3190)
Event Date 11/21/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: regen/rnglc+ ltd 56mm sz 24 catalog#: pt-116056 lot#: k070369.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2016-05242.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the shell was implanted and the surgeon could not free the shell inserter from the shell as it was cold welded into the shell.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.The cup was able to be freed from the inserter.Visual inspection of the inserter shows signs of wear and tear over a service lifetime of approximately 4 years.The front gear is fractured and has two gears/ teeth missing.The square end of the inserter has some debris on it where it was joined to the cup; once separated, there is some evidence of corrosion in the threads.Sem analysis of the fractured surfaces of the front gear of the inserter indicated the fracture occurred due to brittle overload.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIH MOD CUP INSERTER
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6801926
MDR Text Key82981920
Report Number0001825034-2017-06467
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PN/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number31-400600
Device Lot NumberZB130401
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/16/2017
Initial Date FDA Received08/17/2017
Supplement Dates Manufacturer Received05/08/2018
Supplement Dates FDA Received05/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/19/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-