Belt sn (b)(4) was returned and evaluated at the distributor in accordance with procedures recommended by zoll manufacturing corporation.Upon evaluation, the trunk cable connector was cracked and the black main batt wire was open inside the trunk cable.The cracked connector and open black wire cannot be ruled out as a potential contributing factors to the reported service code 107 and insecure connection between the monitor and electrode belt.The root cause of the cracked connector and open wire is excessive force.No adverse event resulted from the defective electrode belt.
|