• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO PROTACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO PROTACK; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 174006
Device Problems Detachment Of Device Component (1104); Physical Resistance (2578); Mechanics Altered (2984)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2017
Event Type  malfunction  
Manufacturer Narrative
User facility listed in initial reporter.(b)(4).This event was originally reported on a quarterly summary report and is being resubmitted per fda request.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopy, the device was not firing properly.After 3 presses on the handle the clamp becomes jammed and several tacks came out at the same time.Tacks fell into the patient cavity but were retrieved.Another device was used in other to complete the case.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.There were no visual or functional abnormalities observed during functional analysis.There were no disengaged components and the handle was able to be actuated.The remaining tacks seated properly in the test media.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Our investigation was unable to determine a root cause or establish a relationship between the device and the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTACK
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
sharon murphy
60 middletown avenue
north haven, CT 06473
2034925267
MDR Report Key6811722
MDR Text Key83313556
Report Number2647580-2017-06190
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier20884521070407
UDI-Public20884521070407
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K090470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model Number174006
Device Catalogue Number174006
Device Lot NumberP7A0163PX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-