Results: the 3penumbra system 3max reperfusion catheter (3maxc) was kinked approximately 5.0 cm from the hub.Conclusion: evaluation of the returned device confirmed that the 3maxc was kinked.This type of damage typically occurs due to improper handling during removal from the packaging.If the device is forcefully retracted out of its packaging hoop at extreme angles, damage such as this may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a thrombectomy procedure, the hospital technologist noticed that a penumbra system 3max reperfusion catheter (3maxc) was kinked upon removal from the packaging.The damaged 3maxc was found prior to use and therefore, was not used in the procedure.
|