Model Number 3186
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Muscle Spasm(s) (1966); Device Overstimulation of Tissue (1991); Pain (1994)
Event Date 03/17/2017
Event Type  Injury  
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 3 of 3.Reference mfr report: 1627487-2017-04712.Reference mfr report: 1627487-2017-04713.It was reported ((b)(6)) the patient experienced a significant spasm while in x-ray.The patient was fully oriented upon examination and scs system diagnostics were normal.The patient alleged receiving overstimulation sensations down both legs with pain shooting up his upper back and arm.The patient also described feeling a warm sensation at ipg site and pain at the lead sites which occurs with stimulation turned on and off.A representative met with the patient for diagnostic testing.All contacts were within normal impedance range and the system was operating as intended.X-rays confirmed the leads were still in proper position to capture targeted pain areas.Reprogramming was unable to resolve the issue.The patient underwent surgical intervention where the entire scs system was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6827327
MDR Text Key83878853
Report Number1627487-2017-04714
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nursing Assistant
Type of Report Initial
Report Date 08/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2017
Device Model Number3186
Device Lot Number5128330
Other Device ID Number05414734401708
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-