STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON #2 PS INSERT 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C
|
Back to Search Results |
|
Catalog Number 5532P209 |
Device Problems
Device Operates Differently Than Expected (2913); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/08/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
During a total knee replacement both femoral and tibial implants were inserted without any problems.While inserting the tibial bearing insert, it would not engage into the tibial tray fully.As a result, the insert was removed and replaced with another, which engaged correctly.
|
|
Manufacturer Narrative
|
An event regarding a seating/locking issue involving a triathlon insert was reported.The event was confirmed by visual inspection.The insert was returned without the locking wire.There is damage on the anterior edge of the device most likely caused during removal of the insert from the baseplate after attempted assembly.There is damage on the posterior aspect of the bearing surface of the device.There is damage present on the anterior outside rim of the bearing surface.There is also damage on the underside of the component on multiple areas.These types of damage are indicative of an obstruction on the posterior aspect of the insert and on the retaining slot of the inferior surface which may have been the reason that the user failed to position the insert correctly on the baseplate.Medical records received and evaluation: not performed as the reported component was not implanted.Dhr review for the reported lot was satisfactory.Chr review for the reported lot confirmed that there are no other similar events reported.Based on the visual inspection of the returned component, the observed damage on the posterior aspect and inferanterior outside rim of the bearing surface of the insert is indicative of an obstruction during attempted seating of the insert onto the baseplate.A capa trend analysis was conducted for the reported failure mode and concluded that seating/locking difficulties may arise from user-related issues associated with the preparation of the joint space prior to the attempted seating and with the technique required to correctly introduce the insert component to the baseplate.No further investigation for this event is required at this time.
|
|
Event Description
|
During a total knee replacement both femoral and tibial implants were inserted without any problems.While inserting the tibial bearing insert, it would not engage into the tibial tray fully.As a result, the insert was removed and replaced with another, which engaged correctly.
|
|
Search Alerts/Recalls
|
|
|