Visual inspection was performed on the returned sample, during which dried blood was visible within the valve.Upon evaluation of the returned sample, it was found to function as intended, and met all of the product specifications.Due to the timing of this event and the manufacture date of the valve, this is believed to be an issue with the forward flow through the duckbill valve.The reported issue was not able to be replicated; therefore, the complaint was not confirmed.The initial failure in negative pressure relief is likely due to dried blood around the negative umbrella valve, causing it to seal shut.Once the seal was broken and pressure was released, it was able to function as intended.There is a known issue currently ongoing with the duckbill valve within the ops valve, that prevents the duckbill from opening within the product specification range.The cause of the issue with the duckbill is due to a process change at the supplier, specifically during the washing/cleaning step; however, the (b)(4).All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
|