• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) Back to Search Results
Model Number DM0010FAA
Device Problem Device Remains Activated (1525)
Patient Problem Perforation (2001)
Event Date 08/03/2017
Event Type  Injury  
Manufacturer Narrative
Report inconclusive.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.This dm0010faa perforator with lot number 051/17 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to monitor this complaint type for trends.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported by the sales representative that the surgeon was drilling burr holes and experienced a dural tear.The neuro resident reported that he felt a sort of a wobbling sensation as he hit the inner table, so he let off the pedal for a moment.When he re-engaged the pedal fully and put pressure on the motor, the tool pushed through the shelf and hit the dura.According to the resident, it did not hit the brain directly, but caused a slight dural tear.The attending neurosurgeon, used gelfoam in the burr hole for hemostasis and moved forward with only a two or three minute delay.It was reported that there was no additional impact to the case.On follow up, it was reported the event occurred during the second burr hole.The first burr hole was made without complications.The patient was recovering from surgery as per normal protocols.The surgical site was the frontal/temporal bone and there were no specific indications of bone pathologies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -390
BR  09980-390
Manufacturer (Section G)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -390
BR   09980-390
Manufacturer Contact
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980--390
77886587
MDR Report Key6836545
MDR Text Key84300520
Report Number1625507-2017-00031
Device Sequence Number1
Product Code HBF
UDI-Device Identifier07898959543081
UDI-Public07898959543081
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/01/2019
Device Model NumberDM0010FAA
Device Catalogue NumberDM0010FAA
Device Lot Number051/17
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age50 YR
-
-