|
Model Number H7493911332250 |
Device Problem
Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/14/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device is a combination product.(b)(4).Promus element,mr,ous 2.50x32mm stent delivery system (sds) was returned for analysis.A visual and microscopic examination of the crimped stent identified stent damage.The distal end of the stent was damaged and stretched in a distal direction over the distal marker band.The crimped stent od(outer diameter) of the undamaged section of the stent was measured and the result was within max crimped stent profile measurement.The balloon was reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the shaft polymer extrusion found a kink in the inner and outer shaft polymer extrusion 40mm distal from the port bond.A visual and microscopic examination found no issues with the tip.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2017.It was reported that shaft kink occurred.The target lesion was located in a coronary artery.A 2.50x32mm promus element drug-eluting stent was advanced to treat the lesion.However, during procedure, the delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed distal stent damage.
|
|
Search Alerts/Recalls
|
|
|