• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/75
Device Problems High impedance (1291); Connection Problem (2900); No Pacing (3268)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/24/2017
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the lead was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During follow-up, high pacing impedance and no pacing output were noted on the left ventricular (lv) lead.The lead was revised and it was noted that the lead was not properly inserted into the lv port of the device header.Once the lead was re-connected to the device, all electrical measurements were normal and in range.The lead and the device remain implanted and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6839794
MDR Text Key84612231
Report Number2017865-2017-07935
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model Number1458Q/75
Device Lot NumberA000037816
Other Device ID Number05414734503334
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2017
Initial Date FDA Received09/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-