• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-Q180
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/28/2017
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.Omsc reviewed the manufacturing history of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available or if the device is returned at a later time, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that in the routine microbiological test by the user facility, the microbes were detected as follows, from samples taken from the subject device.The user facility reported that the subject device was reprocessed manually with peracetic acid disinfectant solution (anioxyde 1000), and they also reported that the subject device was reprocessed according to the instruction manual.There was no report of infection associated with this report.<1st time> - all channels: enterobacter sp.( >100 cfu).<2nd time> - instrument channel: enterobacter sp.( >100 cfu).- suction channel: unspecified microbes ( >100 cfu).
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus (b)(4).(b)(4) sent this device to a third party laboratory for additional microbiological testing.As a result of additional microbiological testing by a third party laboratory, no microbe was detected from samples taken from the instrument channel, suction channel and air/water channel of this device.Therefore, it cleared the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is submitting to correct "device product code".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
katsuaki morita
2951 ishikawa-cho
hachioji-shi, tokyo-to 
426425177
MDR Report Key6840859
MDR Text Key86581836
Report Number8010047-2017-01278
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK100584
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-Q180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/08/2017
Initial Date FDA Received09/04/2017
Supplement Dates Manufacturer Received09/08/2017
02/13/2019
Supplement Dates FDA Received09/26/2017
02/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-