Catalog Number 26-1221
Device Problem Failure to Auto Stop (2938)
Patient Problem Tissue Damage (2104)
Event Date 08/16/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Upon completion of the investigation a follow up report will be filed.
 
Event Description
It was reported that the perforator¿s automatic disengagement mechanism did not work when the surgeon was making the 1st burr hole on (b)(6) 2017.It caused dura mater injury and brain parenchymal damage.The surgery was completed by using new perforator without any further trouble.The surgeon commented that the surgeon applied the perforator vertically to the skull when making the 1st hole.The patient¿s condition is stable now, and there is no critical damage to the patient.No further information was provided by hospital.
 
Manufacturer Narrative
Device was returned for evaluation.A follow up report will be filed upon completion of the investigation.
 
Manufacturer Narrative
Upon completion of the investigation it was noted that the visual inspection utilized the unaided eye.No anomalies were observed.Test performed , unit was found to performed as intended.Test performed, unit completed testing (5 holes) and was found to performed as intended.The dhr was reviewed by and it was verified that there were no anomalies during the manufacturing process.Complaint could not be verified.Unit was found to meet all acceptance criteria.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key6842077
MDR Text Key84700474
Report Number1226348-2017-10659
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberH77316
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received11/29/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-