• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY MAILLEFER LENTULO RA; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY MAILLEFER LENTULO RA; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A002222100100
Device Problems Break (1069); Material Separation (1562)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
While there is no indication that a serious injury resulted in this event, there has been a previous report where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a lentulo ra 21mm separated; the outcome is unknown as of this mdr evaluation.However, there is no indication that injury resulted.
 
Manufacturer Narrative
Multiple unsuccessful attempts were made to obtain the patient outcome.Involved root filler is actually broken in the active part and no material defect was found during analysis of the rupture pattern.No unused product is available for evaluation.The batch number is unknown, dhr cannot be reviewed.Root causes are not identified.We will track this kind of event and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENTULO RA
Type of Device
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ  1338
Manufacturer (Section G)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ   1338
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6853347
MDR Text Key86622386
Report Number8031010-2017-00103
Device Sequence Number1
Product Code EKS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA002222100100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/11/2017
Initial Date FDA Received09/08/2017
Supplement Dates Manufacturer Received10/24/2017
Supplement Dates FDA Received11/20/2017
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-