Device evaluation summary: the everflex entrust stent delivery system, (sds), was received for evaluation.The entrust sds was received loaded on a 0.035¿ compatible guidewire.A portion of the stent was exposed.The distal end of the exposed stent did not have any of the stent radiopaque marker hoops.The distal end of the exposed stent exhibited tearing of the stent struts.The proximal end of the stent was located using back lighting.Approximately 7mm of the 150mm stent were not returned with the entrust stent deployment system.The blue isolation sheath was moved distally exposing the inner guidewire lumen/pusher.Dried sanguine residue was noted along the length of the freshly exposed inner guidewire lumen/pusher.It was noted that the outer sheath was moving along with the isolation sheath.It was also noted that the guidewire was able to be removed from the proximal hub luer lock with ease after the isolation sheath and outer sheath were moved distally.The blue strain relief/isolation sheath was skived off the gold isolation sheath until the blue strain relief could be advanced proximally off the gold isolation sheath and silver outer sheath.It was noted that the pull cable was still attached to the outer sheath.It was also noted that the gold isolation sheath and silver outer sheath could not move independent of each other.Dry sanguine residue was note on the exterior of the exposed silver outer sheath.Back lighting was used to locate the distal end of the inner guidewire lumen/pusher.The distal end of the inner guidewire lumen/pusher was located approximately 131cm from the distal end of the proximal hub.The catheter section of entrust stent delivery system was allowed to soak over 72 hours to re-hydrate the dry sanguine residue in order to flush it from the delivery system.The guidewire returned was re-loaded into the entrust stent delivery system.The guidewire returned with the device exhibits drag witness marks from where it interacted with the stent.Stands of skived ribbons from the returned guidewire are visible on the stent struts.A mass of skived ribbons from the returned guidewire was noted on the guidewire.The stent was able to be deployed after rehydration of the sanguine residue by pinning the guidewire loaded inner and manually pulling the outer sheath.At the distal end of the inner guidewire lumen/pusher was a mass of sanguine residue.Transparent strands of polymer were observed on the exterior of the inner guidewire lumen/pusher.Approximately 143mm of the stent of the 150mm were returned.If information is provided in the future, a supplemental report will be issued.
|