• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DISTAFLO BYPASS GRAFT; EPTFE VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DISTAFLO BYPASS GRAFT; EPTFE VASCULAR GRAFT Back to Search Results
Model Number DFX8006SC
Device Problem Torn Material (3024)
Patient Problem No Patient Involvement (2645)
Event Date 08/11/2017
Event Type  malfunction  
Manufacturer Narrative
Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Visual/microscopic inspection: as the device was not returned, an inspection could not be performed.Functional/performance evaluation: as the device was not returned, an evaluation could not be performed.Medical records review: no medical records have been made available to the manufacturer.Image/photo review: no medical images have been made available to the manufacturer.Conclusion: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive for the alleged crack/torn material as no objective evidence has been provided to confirm any alleged deficiency with the device.The definitive root cause could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu states: precautions: when removing the external spiral support (beading) of the distaflo® graft, the beading must be removed slowly and at a 90° angle to the graft.Rapid unwinding and /or removal at less than a 90° angle may result in graft damage.Do not use surgical blades or sharp, pointed instruments to remove the beading as this may damage the graft wall.If damage occurs, that segment of the graft should not be used.Aggressive and/or excessive graft manipulation when tunneling, or placement within a too tight or too small tunnel, may lead to separation of the spiral beading and/or graft breakage.The distal anastomosis should be made after tunneling or suture disruption can occur.Do not pass the cuff portion (distal end) of the distaflo bypass graft through a tunneler sheath or the tissue tunnel, as this could lead to separation of the spiral beading and/or graft breakage.Distaflo bypass grafts do not stretch (are non-elastic) in the longitudinal direction.The correct graft length for each procedure must be determined by considering the patient's body weight, posture, and the range of motions across the anatomical area of graft implantation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that during preparation, a vascular graft was allegedly cracked at the graft cuff.A new vascular graft was used to complete the procedure.There was no reported patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTAFLO BYPASS GRAFT
Type of Device
EPTFE VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key6853789
MDR Text Key85760800
Report Number2020394-2017-01103
Device Sequence Number1
Product Code DSY
UDI-Device Identifier00801741023675
UDI-Public(01)00801741023675
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K983861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/10/2021
Device Model NumberDFX8006SC
Device Catalogue NumberDFX8006SC
Device Lot NumberVTAZ0223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-