Model Number ZM-520PA
Device Problems Overheating of Device (1437); Smoking (1585); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/14/2017
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer reported that the transmitter overheated and was smoking.The battery contacts were burned and melted.There was no other physical damage or fluid intrusion reported.The device was in use on a patient, however no patient injury was reported.They were provided with an exchanged device and sent the failed device in to nihon kohden.It is currently awaiting evaluation.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The biomedical engineer reported that the transmitter overheated and was smoking.The battery contacts were burned and melted.
 
Manufacturer Narrative
The biomedical engineer reported that the transmitter overheated and was smoking.The battery contacts were burned and melted.There was no other physical damage or fluid intrusion reported.The device was in use on a patient, however no patient injury was reported.They were provided with an exchanged device and sent the failed device in to nihon kohden for evaluation.It was determined in the evaluation that the cause of the unit overheating was due to the customer incorrectly installing the battery in to the unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-520PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key6855773
MDR Text Key85787801
Report Number8030229-2017-00297
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921115077
UDI-Public04931921115077
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-520PA
Device Catalogue NumberZM-520PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Report Sent to FDA09/11/2017
Distributor Facility Aware Date08/14/2017
Device Age63 MO
Event Location Hospital
Date Report to Manufacturer09/11/2017
Date Manufacturer Received09/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse