Model Number 3100A |
Device Problems
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/15/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Vyaire medical complaint number (b)(4).At this time, vyaire medical has not received the suspect device/component from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that the ventilator was not running at some point, while in use on a patient.The unit stayed pressurized.The unit was swapped out and there was no patient compromise.The customer put a test patient circuit on the unit and let the unit run for 8 hours without any issues.
|
|
Manufacturer Narrative
|
A vyaire medical field service representative (fsr) evaluated the device onsite.The fsr found that the pr3 pressure regulator was set too high and the ventilator was hitting the 47 cmh2o limit.The fsr also found that the pr3 was drifting slighty and replaced it.The fsr found that the frequency panel meter was out of specification.The fsr replaced the panel meter and calibrated the pressure transducer, pneumatics functions, driver controller, and driver displacement indicator.The device passed the alarms test, performance check, and patient circuit calibration.The fsr completed the operational verification procedure (ovp).The device meets manufacturer specifications.
|
|
Manufacturer Narrative
|
Results of investigation: the vyaire medical failure analysis laboratory received the suspect components, a 3100a panel meter and pr3 regulator.An evaluation of the components did not duplicate the reported issue.The panel meter and pr3 regulator meet manufacturer's specifications.
|
|