Model Number CI-1400-01 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/24/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi number: na.
|
|
Event Description
|
The recipient reportedly experienced intermittencies and loss of lock.External equipment was exchanged and programming adjustments were made, however, the issue is not resolved.Revision surgery is under consideration.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.Legal proceedings have prohibited the testing and failure analysis of the explanted device.As a result, no conclusion can be drawn at this time.If the legal proceedings allow for the analysis to be completed, the issue will be re-opened and the results of the analysis will be reported.The external visual inspection revealed the electrode was severed at the flange and near the array prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed electrode damage at the flange near the fantail.This is believed to have occurred during revision surgery.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The device failed the residual gas analysis test.This is the final report.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.This is the final report.
|
|
Manufacturer Narrative
|
The external visual inspection revealed tool marks on the top and bottom covers and along the seam weld, and the electrode was severed at the flange and near the array prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection confirmed electrode damage at the flange near the fantail.This is believed to have occurred during revision surgery.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The device failed the residual gas analysis test.The scanning electron microscopy analysis revealed tool mark damage on the seam weld.This is believed to have occurred during revision surgery.The internal visual inspection noted silver migration on some electrical components.The residual gas analysis result was above the nitrogen limit.This result is due to a gross leak from damage to the seam weld, which is believed to have occurred during revision surgery.This device had moisture that exceeded the residual gas analysis test limit.Leak testing of the feedthru assembly did not reveal any leaks.Corrective actions have been implemented.This is the final report.
|
|
Search Alerts/Recalls
|