Model Number 401180
Device Problems Impedance Problem (2950); No Pacing (3268)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/08/2017
Event Type  malfunction  
Event Description
Ous mdr - during a follow up visit this lv lead was not pacing and the impedance was changed.The lead was explanted and replaced.No adverse patient events were reported.
 
Manufacturer Narrative
Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the lead was scrutinized, including a visual and electrical inspection.In the course of the analysis, no deviations were noted.The lead proved to be without fault throughout its inspection.In particular, the values of the parameters measured during the electrical analysis were within the technical specifications.Further analysis of the lead did not reveal any irregularity that might have contributed to the reported clinical observation.The manufacturing process for this lead was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.Biotronik monitors the post-market performance of its products closely in order to identify any trends that may reveal over time a potential manufacturing or design issue.The historic performance of the product involved in the current case does not at this point reveal any such trend.For this reason we encourage close dialogue between clinicians and our product experts in order to explore all options to minimize any such occurrences in future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP S-85
Type of Device
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin D-123 59
GM  D-12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key6874750
MDR Text Key86716793
Report Number1028232-2017-03320
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number401180
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/12/2017
Initial Date FDA Received09/18/2017
Supplement Dates Manufacturer Received10/16/2017
Supplement Dates FDA Received10/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-