Model Number 595000-001
Device Problem Device Displays Incorrect Message (2591)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 09/03/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited intermittent onboard battery alarms, the freedom driver continued to perform its life-sustaining functions.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited intermittent onboard battery alarms while supporting a patient.The customer also reported that the patient was subsequently switched to the backup freedom driver.There was no reported adverse patient impact.
 
Manufacturer Narrative
The freedom driver was returned to syncardia for evaluation.The customer-reported intermittent onboard battery alarms were not able to be confirmed, as intermittent alarms are not recorded in the driver's alarm history.Intermittent onboard battery alarms were not able to be reproduced as part of normal driver operation.Moreover, onboard batteries in-use with the driver during the customer experience were not returned and were therefore not included in this investigation.The driver functioned as intended, including exhibiting an alarm when a low charge onboard battery was inserted or an onboard battery was missing.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The customer, a syncardia certified hospital, reported that the freedom driver exhibited intermittent onboard battery alarms while supporting a patient.The customer also reported that the patient was subsequently switched to the backup freedom driver.There was no reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6876079
MDR Text Key87855677
Report Number3003761017-2017-00178
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 09/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/07/2017
Initial Date FDA Received09/19/2017
Supplement Dates Manufacturer Received09/22/2017
Supplement Dates FDA Received10/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age43 YR
-
-