Model Number 595000-001 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/03/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited intermittent onboard battery alarms, the freedom driver continued to perform its life-sustaining functions.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the freedom driver exhibited intermittent onboard battery alarms while supporting a patient.The customer also reported that the patient was subsequently switched to the backup freedom driver.There was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The freedom driver was returned to syncardia for evaluation.The customer-reported intermittent onboard battery alarms were not able to be confirmed, as intermittent alarms are not recorded in the driver's alarm history.Intermittent onboard battery alarms were not able to be reproduced as part of normal driver operation.Moreover, onboard batteries in-use with the driver during the customer experience were not returned and were therefore not included in this investigation.The driver functioned as intended, including exhibiting an alarm when a low charge onboard battery was inserted or an onboard battery was missing.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the freedom driver exhibited intermittent onboard battery alarms while supporting a patient.The customer also reported that the patient was subsequently switched to the backup freedom driver.There was no reported adverse patient impact.
|
|