Model Number 595000-001
Device Problems Break (1069); Physical Property Issue (3008)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 08/31/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Event Description
The freedom driver was not supporting a patient at the time of the reported event.The customer, a syncardia certified hospital, reported that the patient's backup freedom driver had a damaged driveline strain relief.
 
Manufacturer Narrative
The freedom driver was returned to syncardia for evaluation.The customer-reported issue of a frayed strain relief was confirmed through visual inspection of the driver.The inspection revealed markings on the driveline under the damaged part of the strain relief.It is likely that the strain relief was struck by something, perhaps when setting the driver down, as there were marks on the drivelines on the underside of the driver.The root cause of the customer-reported issue could not be conclusively determined.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The freedom driver was not supporting a patient at the time of the reported event.The customer, a syncardia certified hospital, reported that the patient's backup freedom driver had a damaged driveline strain relief.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6876083
MDR Text Key87740074
Report Number3003761017-2017-00172
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age21 YR
-
-