• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. CANNULATED POLYAXIAL SCREW 4.5MM X 45MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. CANNULATED POLYAXIAL SCREW 4.5MM X 45MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3505-4545
Device Problem Disassembly (1168)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2017-00504 thru 3012447612-2017-00506.
 
Event Description
It was reported that three pedicle screws disassembled during surgery.While a previously installed construct was being loosened, the tulips detached from the threaded shafts.They were removed and replaced with alternative screws.There were no reports of patient injury associated with this event.This is report one of three for this event.
 
Manufacturer Narrative
The returned screw was examined.The screw components were found to have disassembled due to deformation on the swivel rings, which was caused when the screw was locked into position, then unlocked and repositioned during the procedure.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED POLYAXIAL SCREW 4.5MM X 45MM
Type of Device
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key6877013
MDR Text Key87337776
Report Number3012447612-2017-00504
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK132884
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3505-4545
Device Lot Number84WW
Other Device ID Number(01)00889024335592
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/21/2017
Initial Date FDA Received09/19/2017
Supplement Dates Manufacturer Received02/28/2018
Supplement Dates FDA Received03/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-