MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/16/2017
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the sheath, 4fc12 was returned and analyzed.Visual inspection of the sheath showed the shaft was kinked/twisted at 1.1115 inches, 1.6845 inches and 2.4265 inches from the tip.Functional testing of the sheath showed the deflection was out of plane.Dissection did not show any breakage of the pull wire.In conclusion, the deflection issue was confirmed through testing.The sheath failed the return product inspection due to multiple kinks/twists in the shaft.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the sheath was not steerable.The sheath was replaced and the case was completed with cryo.No patient complications have been reported as a result of this event.The sheath subsequently tested out of specification per the manufacturer¿s investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6879878
MDR Text Key87862459
Report Number3002648230-2017-00523
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/26/2019
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number92439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received09/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-