• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40Q
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332); No Device Output (1435)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2017
Event Type  Injury  
Event Description
It was reported that the patient presented in-clinic on (b)(6) 2017 after feeling what he thought was a vibratory alert.The implantable cardioverter defibrillator was not able to be interrogated.An electrocardiogram was performed and showed a complete loss of output.Premature battery depletion was suspected.The device was explanted and replaced.The patient was stable post procedure.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with lithium (li) cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6880169
MDR Text Key86770805
Report Number2017865-2017-32062
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2015
Device Model NumberCD3357-40Q
Device Catalogue NumberCD3357-40Q
Device Lot Number4341802
Other Device ID Number05414734508216
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0116-2017
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
-
-