• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT 5300

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT 5300 Back to Search Results
Model Number 1790103
Device Problems Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/22/2017
Event Type  malfunction  
Manufacturer Narrative
The affected device was replaced at the facility.The defective parts will be evaluated once they have been returned.This failure mode has been previously investigated and it was found that the damage to the painted surfaces usually does not develop suddenly, but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The surface coating process will be updated to include additional testing and verification.A follow-up report will be submitted if more information becomes available.
 
Event Description
The powder coat paint was found to be chipping from a trumpf medical surgical light lamp head.No injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT 5300
Type of Device
TRULIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key6891011
MDR Text Key87868491
Report Number9681407-2017-00031
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K102758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial
Report Date 06/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1790103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/22/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/24/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-