• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE; VALVE SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE; VALVE SYRINGE Back to Search Results
Catalog Number 0035280
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the user found black material inside the device prior to use.
 
Manufacturer Narrative
Received 1 unopened syringe bulb.The reported event was unconfirmed.During the visual inspection, the sample was reviewed by the following requirement specified in the inspection procedure: loose foreign matter or embedded shall not exceed an aggregate total of 0.6mm² or 1/16¿ per tappi dirt estimation chart and no obvious defects were noted in the sample.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "irrigation syringe after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Caution: federal (usa) law restricts this device to sale by or on the order of a physician.Do not resterilize.This is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." (b)(4).
 
Event Description
It was reported that the user found black material inside the device prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE
Type of Device
VALVE SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6891040
MDR Text Key88258871
Report Number1018233-2017-04951
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier00801741049255
UDI-Public(01)00801741049255
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Catalogue Number0035280
Device Lot NumberNGBR0903
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/04/2017
Initial Date FDA Received09/25/2017
Supplement Dates Manufacturer Received10/23/2017
Supplement Dates FDA Received10/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-