Model Number 0.850.0022
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 07/26/2017
Event Type  malfunction  
Manufacturer Narrative
The initial report did not indicate that the suspect device experienced a malfunction that would be a reportable event.After an initial visual inspection of the device, it was determined that the event that occurred is reportable and an evaluation was conducted.Upon visual inspection it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the upper battery pack between the cells and printed circuit board.There was a blackening on the top of the upper battery pack and its printed circuit board (pcb).Cells 5 and 6 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes the investigation.
 
Event Description
It was reported that the office is receiving a system failure message with every battery that they put on the unit.There was no report of injuries, or patient or user involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO 2
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6901728
MDR Text Key87746354
Report Number1017522-2017-00053
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0022
Device Catalogue Number0.850.0022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/26/2017
Initial Date FDA Received09/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-