The initial report did not indicate that the suspect device experienced a malfunction that would be a reportable event.After an initial visual inspection of the device, it was determined that the event that occurred is reportable and an evaluation was conducted.Upon visual inspection it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the upper battery pack between the cells and printed circuit board.There was a blackening on the top of the upper battery pack and its printed circuit board (pcb).Cells 5 and 6 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes the investigation.
|