Product analysis: upon receipt at medtronic¿s quality laboratory, the valve showed evidence of blood contact.White multifilament suture was found attached to the sewing cloth adjacent to the left cusp.A cut was observed through the sewing cloth adjacent to the left cusp, which appeared to be from a sharp object.The damage to the cloth extended up to the outflow rail.All leaflets were in open position.All leaflets were slightly stiff but flexible.All commissures were intact.Tears on the left cusp lunula were observed.A tear along the belly of the non-coronary cusp was observed.The device was visually inspected by manufacturing quality.Per the manufacturing engineers, the type of damage found on the left cusp and non-coronary cusp can be detected during manufacturing.Conclusion: the device history record was also reviewed and shows that this valve met all manufacturing specifications for product released to distribution.Exactly what caused (when and how) the tissue damage cannot be determined; however during 100% inspection of the final assembly at medtronic, the tissue was not damaged.Based on the received information and analysis results of the returned device, the clinical observations, tear, was confirmed.The cause of the tear was probably related to the accidental damage during preparation / implant with a sharp instrument.If information is provided in the future, a supplemental report will be issued.
|