• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY Back to Search Results
Catalog Number 295025-001
Device Problems Battery Problem (2885); Charging Problem (2892); Power Problem (3010)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom onboard battery was not supporting a patient.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce 3980 initial.
 
Event Description
The freedom onboard battery was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom onboard battery did not hold a charge.
 
Manufacturer Narrative
The battery was returned in an over-discharged state requiring use of the built-in zero-voltage-charge (zvchg) recovery utility.The over-discharge condition would have caused the customer-reported issue of being unable to hold a charge.The root cause for the over-discharged condition was that the battery was allowed to become discharged and never received a maintenance charge.The battery was successfully recovered by use of the built-in zvchg recover utility and was tested in accordance with the current evaluation process where it successfully passed all sections.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6907594
MDR Text Key89695544
Report Number3003761017-2017-00191
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2017
Initial Date FDA Received10/02/2017
Supplement Dates Manufacturer Received09/21/2017
Supplement Dates FDA Received09/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-