Catalog Number 295025-001 |
Device Problems
Battery Problem (2885); Charging Problem (2892); Power Problem (3010)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom onboard battery was not supporting a patient.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce 3980 initial.
|
|
Event Description
|
The freedom onboard battery was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom onboard battery did not hold a charge.
|
|
Manufacturer Narrative
|
The battery was returned in an over-discharged state requiring use of the built-in zero-voltage-charge (zvchg) recovery utility.The over-discharge condition would have caused the customer-reported issue of being unable to hold a charge.The root cause for the over-discharged condition was that the battery was allowed to become discharged and never received a maintenance charge.The battery was successfully recovered by use of the built-in zvchg recover utility and was tested in accordance with the current evaluation process where it successfully passed all sections.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|