MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 2088TC/52
Device Problems Bent (1059); Difficult to Insert (1316); Kinked (1339); Under-Sensing (1661); Low impedance (2285); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/07/2017
Event Type  malfunction  
Event Description
It was reported that the asymptomatic patient presented for a normal pulse generator upgrade procedure.After removing and replacing the device, the right atrial lead exhibited a loss of atrial pacing and low impedance.The physician had difficulty inserting a stylet into the lead and experienced difficulty retracting the helix.The lead then dislodged and was removed and replaced.Upon removal, it was noted that the stylet had punctured the insulation of the lead and that there was a kink in the lead.The procedure was completed successfully and patient was stable following the procedure.
 
Manufacturer Narrative
(b)(4).Analysis was normal.Anomalies found were consistent with those occuring during surgical procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6909744
MDR Text Key88020031
Report Number2017865-2017-08706
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model Number2088TC/52
Device Lot NumberA000037387
Other Device ID Number05414734502887
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age90 YR
-
-