The sample tube (bd 1.8ml double walled tube) used for analysis was unacceptable for use on the analyzer.The sample probe was hitting the side of the tube while aspirating.No analyzer deficiencies were identified.Per the instructions for use (ifu) for the analyzer, the only double walled tubes to be used on the analyzer are the vacutainer plus plastic citrate tube, 2.7 ml or the vacuette sandwich coagulation tube, 3.0 ml/3.5ml.The ifu specifically identifies the tube used as not being acceptable for use on the analyzer.
|