ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Information (3190)
Event Date 08/29/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (loose connection at monitor) has been confirmed.Upon investigation the trunk cable connector was physically damaged and the electrode belt was unable to securely connect to a monitor.The root cause for the damaged connector was physical abuse.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor reported that a patient's electrode belt connector was coming loose from the monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
allison petzold
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key6916841
MDR Text Key89465480
Report Number3008642652-2017-08361
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005012
UDI-Public00855778005012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2017
Was the Report Sent to FDA? No
Date Manufacturer Received09/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-