Model Number G141 |
Device Problems
High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
|
Patient Problems
Atrial Flutter (1730); No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/12/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
Boston scientific received information that atrial tachycardia response (atr) events were observed recently with this system and the atrial channel appears to demonstrate noise due to interaction with rate response trend (rrt) feature.A review of the daily measurements noted an increase in pacing impedance measurements.The patient was going to be brought into the office for further troubleshooting and to possible program off rrt.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Additional information was received stating this patient is in chronic atrial fibrillation (af) and noise and oversensing continued to result in atr episodes.A boston scientific technical services consultant discussed the clinical observations and again recommended programming off the respiratory rate trend (rrt) feature.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|