• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON AQUATHERM HEATER 091; HEATER, BREATHING SYSTEM W/WO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON AQUATHERM HEATER 091; HEATER, BREATHING SYSTEM W/WO Back to Search Results
Catalog Number 048-91
Device Problems Insufficient Heating (1287); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device was not returned to the manufacturer at the time of this report.A device history record investigation did not show issues related to this complaint.A record assessment of revision of (b)(4) was performed and the failure mode is already included.No update is required.Customer complaint cannot be confirmed based on the information received.Root cause is unknown.No corrective actions can be established at this time.It is necessary to receive the device sample to perform a proper and thorough investigation to confirm the alleged defect reported, determine a root cause, and establish corrective actions.If the device sample becomes available at a later date, this report will be updated accordingly.
 
Event Description
Customer complaint alleges the device is not heating.Alleged issue reported as detected prior to patient use during functional testing.There was no report of patient harm or delay in treatment.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was performed; however, a current test could not be performed due to an open circuit.It is possible that the open circuit was due to overheating of the unit.Based on the investigation performed, the reported complaint was confirmed.All aquatherm heaters are 100% inspected during manufacturing; therefore, a defect of this type would be detected prior to release from the manufacturing facility.It was determined that operational context caused or contributed to the reported defect.
 
Event Description
Customer complaint alleges the device is not heating.Alleged issue reported as detected prior to patient use during functional testing.There was no report of patient harm or delay in treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON AQUATHERM HEATER 091
Type of Device
HEATER, BREATHING SYSTEM W/WO
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key6937552
MDR Text Key90046459
Report Number3003898360-2017-01041
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number048-91
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/25/2017
Initial Date FDA Received10/10/2017
Supplement Dates Manufacturer Received11/16/2017
Supplement Dates FDA Received11/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-