On (b)(6) 2017, a (b)(6) female was involved in an incident with the mayfield modified skull clamp.The patient was prepped for surgery.The swivel lock of the skull clamp was in the locked position, but it still moved and injured the patient.An unspecified revision/medical intervention was required.There was a delay in surgery due to product problem for 10 minutes.Request for additional information was sent.
|
Investigation completed 10./17/2017.With respect to the returned unit it has passed all specific functional testing requirements, except for the lock having rotational movement when unit is not under pressure, this would not have caused a slippage; when unit is properly positioned and put under pressure unit would not have slipped.General maintenance and cleaning required.Device history record reviewed for sn (b)(4) work order /(b)(4) lot/ 154 a total of (b)(4)were manufactured on 5/20/2015 show no abnormalities related to the reported failure.The devices manufactured during this period passed all required inspection points.No manufacturing or design related trend has been identified.Root cause is undetermined at this time.Complaint not confirmed.
|