ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 32MM C; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467); Fitting Problem (2183); Device-Device Incompatibility (2919)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2015
Event Type  malfunction  
Manufacturer Narrative
Zimmer biomet complaint-(b)(4).Reported event was unable to be confirmed due to limited information received from the customer.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.This report is being submitted late as it has been identified in remediation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
Liner would not seat properly.Subsequently, the liner was removed and another liner was used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LNR 32MM C
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6942445
MDR Text Key90302124
Report Number0001825034-2017-08427
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2019
Device Model NumberN/A
Device Catalogue Number010000730
Device Lot Number3400221
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/07/2015
Initial Date FDA Received10/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age68 YR
-
-