• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40C
Device Problem Device Contamination with Body Fluid (2317)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/06/2017
Event Type  malfunction  
Event Description
During lead replacement, blood entered the lead port in the device.The physician decided to explant the device and replace it.The patient was in stable condition.
 
Manufacturer Narrative
The contamination of the header could not be confirmed as inspection revealed no fluid in the header.Interrogation of the device revealed the device was above the elective replacement indicator when received.Pacing, sensing, impedance, high voltage (hv) output, and the patient notifier were tested and no anomaly was detected.The cause of the field event remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6945539
MDR Text Key89597396
Report Number2938836-2017-32776
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2017
Device Model NumberCD3371-40C
Device Lot NumberP000019170
Other Device ID Number05414734508384
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/06/2017
Initial Date FDA Received10/12/2017
Supplement Dates Manufacturer Received12/07/2017
Supplement Dates FDA Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/10/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-