• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 3L RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 3L RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX15RW30C
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 09/25/2017
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion.(b)(4).Results code: results pending completion of evaluation.Conclusions code: conclusion not yet available-evaluation in progress.
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the yellow cap on top of cardiotomy in the venous inlet was broken off and the internal portion was still in the port.No patient involvement.Product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: (b)(4).Visual inspection was performed on the actual sample upon receipt, during which it was found that one of the non-vented yellow caps of the curved venous inlet port was broken.The stem of the cap still remained within the luer port and the luer port had a slight chip along the top portion.Additionally, the reservoir housing had large cracks on the reservoir lid surrounding the venous inlet port location.A retention sample from the same product code and lot number combination was visually inspected and found to have no damages or anomalies, specifically with the reservoir lid and the caps on the curved venous inlet port.All reservoirs are 100% visually inspected at several points in the production process.The packaging of the product also ensures protection against damage to the caps and reservoir.It is likely that the cap and reservoir lid were damaged by a shock force at some point during the handling of the product; however, when or how this shock force was applied was not able to be determined.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda.(b)(4).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX15RWC W/ 3L RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key6947022
MDR Text Key90239517
Report Number1124841-2017-00206
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model Number3CX*FX15RW30C
Device Catalogue NumberN/A
Device Lot NumberVA02
Other Device ID Number(01)00699753450783
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/26/2017
Initial Date FDA Received10/12/2017
Supplement Dates Manufacturer Received10/11/2017
11/22/2017
Supplement Dates FDA Received10/26/2017
11/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-